Flexible Endoscopes FMEA

by 데몬웍스 2016. 10. 4.
반응형

 

Flexible Endoscopes

 

     No.1

  • Reported Issue : Flickering/distorted Image
  • Common Cause : Camera / CCD assembly damaged due to corrosion from fluid invasion

  • Effects : Delayed / cancelled procedure while alternative scope is made available potential increase in anesthesia time for patient

  • Repair : Repair or replace camera

  • Resolution : Educate staff on proper leak testing procedures and document in standard operating procedures.

 

 

     No.2

  • Reported Issue : Inability to adjust
  • Common Cause : Stretched angulation cables/compressed insertion tube

  • Effects : Physician dissatisfaction, poor articulation and control response time

  • Repair : Adjust angulation cables / replace insertion tube and inner components

  • Resolution : Preventative maintenance / regular inspection cycle.

 

     No.3

  • Reported Issue : Tight Controls
  • Common Cause : Previous fluids Invasion / corrosion

  • Effects : Physician dissatisfaction and reduced articulation

  • Repair : Rebuild angulation system

  • Resolution : Educate staff on proper leak testing procedures and document in standard operating procedures.

   

     No.4

  • Reported Issue : Video button not responding
  • Common Cause : Fluid invasion
  • Effects : Delayed / cancelled procedure while alternative scope is made available Potential increase
    in anesthesia time for patient

  • Repair : Replace video switch

  • Resolution : Educate staff on proper leak testing procedures and document in standard operating procedures.