Flexible Endoscopes
No.1
- Reported Issue : Flickering/distorted Image
-
Common Cause : Camera / CCD assembly damaged due to corrosion from fluid invasion
-
Effects : Delayed / cancelled procedure while alternative scope is made available potential increase in anesthesia time for patient
-
Repair : Repair or replace camera
-
Resolution : Educate staff on proper leak testing procedures and document in standard operating procedures.
No.2
- Reported Issue : Inability to adjust
-
Common Cause : Stretched angulation cables/compressed insertion tube
-
Effects : Physician dissatisfaction, poor articulation and control response time
-
Repair : Adjust angulation cables / replace insertion tube and inner components
-
Resolution : Preventative maintenance / regular inspection cycle.
No.3
- Reported Issue : Tight Controls
-
Common Cause : Previous fluids Invasion / corrosion
-
Effects : Physician dissatisfaction and reduced articulation
-
Repair : Rebuild angulation system
-
Resolution : Educate staff on proper leak testing procedures and document in standard operating procedures.
No.4
- Reported Issue : Video button not responding
- Common Cause : Fluid invasion
-
Effects : Delayed / cancelled procedure while alternative scope is made available Potential increase
in anesthesia time for patient -
Repair : Replace video switch
-
Resolution : Educate staff on proper leak testing procedures and document in standard operating procedures.